Our Products
STERI Scrub Sinks have complete SS304 construction and are exceptionally designed to reduce water splashes on user. these sinks are customizable and are the right fit for any operating room. They are available with one, two or three stations. They also come with option of Hands-free operation via proximity sensor assuring safety in all surgical environments.
FEATURES
•Equipped with non-aerating laminar flow hot and cold water faucet: delivering a consistent stream of non-splash water
•Proximity Sensor System
•Inbuilt Water Warming System with Pressure Pump to ensure availability for hot and cold water through automatic solenoid valves
•Knee/Foot Operated Soap Dispenser
•Stainless Steel 304 (16 SWG) construction with continuous welded compartments
•Available models: 1, 2, or 3 station.
OPTIONAL ACCESSORIES
•Wall Mounted Model
•Fully Automatic Water and Soap Dispensing Unit
•Optional Digital Timer
•UV sterilization of water or/and 2 Micron Prefiltration
•Automatic Thermostatic Mixture for centralized hot water
•Eye Wash Station
•Hand held spare with flexible holes and on/off control lever on the handle
FEATURES:
FEATURES:
The frame is mounted on 4 nos heavy duty swivelling castors for easy movement in narrow passages.
FEATURES
•Construction: SS304(16 SWG)
•Enclosure: SS304 (Customization of material is available )
•User friendly with High-performance transducers
•Using upto 45 khz frequency for effective cleaning with low noise
•Ultrasonic Generator is made of advanced latest MOSFET/IGBT based SMPS
•Workable on 220/230 Volts AC Supply
•Available with Timer and Digital Temperature Controller
•Accessories- Lid, Basket and Power Cord
DESIGN AND FEATURES:
FEATURES:
FEATURES:
CSSD rack all sides is open.
Stainless-steel is equipped with bullet type feet.
5 Shelves are provided with vents for air circulation.
Shelves are made of 1.6mm thickness stainless steel 304.
Frame is made of 25x25 square pipe & 1.6mm thickness stainless steel 304.
The equipment is smooth finished without any sharp edges and crevices.
The trolley is also equipped with mobile castors which allow it to be easily moved around a hospital wing.
The STERI Washer Disinfector has a convenient front-loading design equipped with automatic sliding door/s and is best suited for surgical and sterilization departments. Having volumetric capacity of 275 L is more than enough to meet your everyday need. It can be loaded with large variety of specially designed washing trolleys and accessories for surgical instruments, tubular, laparoscopy, ophthalmology, containers, anesthesia and more.
FEATURES
Stainless Steel Construction
Fabrication of complete structure and piping from non-corrosive high quality AISI 304 stainless steel with argon arc welding.
Automatic Enzyme & Detergent Dozing
Equipped with dual dosing pumps with flow regulator and level control for acid/enzymatic and alkaline detergents (2 additional pumps optional)
Advanced Touch Control
8 Preset Decontamination Process Cycles
Large, high-resolution, coloured touch screen gives an easy-to-read overview of real-time Phase, Time and Temperature readings for each selected program.
Automatic Sliding Door
Microprocessor controlled vertical sliding door/s with Silicone gasket for airtight sealing (Available in Single & double door configuration)
Air Drying Prefilter with HEPA H14 Filter
The user is able to control the temperature and time of the drying air within the range of 60° to 120° C
Inbuilt Heating System
Capable of heating water to a temperature of 90 degree Celsius for higher cleaning efficiency
Advanced Features
• 7 Ultra High-Pressure Spray nozzle system for deep rinsing
• Water detection and level sensor system
• Minimal installation needed
The equipment is smooth finished without any sharp edges and crevices.
4 Multicolour Bio Medical waste Bin Trolley.
The bins are also color-coded to help identify the type of waste that should be disposed of in each bin.
Process Control and Other Features:
1. AUTOMATIC CONTROL SYSTEM:
The Machine is equipped with automatic control system.
2. AUTOMATIC PROXIMITY DETECTION SYSTEM:
The unit is provided with advanced Proximity detection system that can sense the presence and allows controlled warm water to flow through electrically controlled solenoid valves and shuts off as soon as person is away from sensor.
3. MANUAL ELBOW OPERATED TAB:
In case there is any unavailability of electricity for automatic controls, elbow operated taps are provided for washing.
4. FOOT OPERATED SWITCH:
Flow & quantity of liquid soap / disinfectant is controlled with the help of foot operated switch.
5. GEYSER SYSTEM:
The machine is equipped with the geyser system.
6. WATER PUMP:
The machine is equipped with the water pump.
7. ELECTRIC SOLENOID VALVES:
Scrub Station is equipped with branded electric solenoid valves for automatic process.
8. AIR GUN SYSTEM:
Scrub station is fitted with the air gun system for drying.
9. WASHER JET:
Scrub Station is equipped with washer jet for washing. & cleaning.
10. EUROPEAN STYLE FULLY COVERED:
Scrub station is fully covered design with panelling Front, back and both the sides RHS & LHS.
11. WATER TANK WITH ACCESSORIES:
The unit also consists of a SS 304 water tank, water heating system and water pump along with thermostat for heater control. This ensures that water is always available irrespective of low pressure.
12. SPECIAL DESIGNED:
The complete unit is made of SS304 (16 gauge) having rounded corners and special flexible extendable washing spout is provided for easy cleaning.
Ethylene Oxide (EO or ETO) Gas is a crucial method for sterilizing delicate items like plastics, optics, and electronics, which are sensitive to high temperatures and radiation. Operating within the range of 30°C to 60°C, with humidity levels above 30% and gas concentrations between 200 and 800 mg/l, this treatment typically spans a minimum of three hours. Renowned for its exceptional penetration abilities, ETO effectively eliminates viruses, bacteria, fungi, and even spores, rendering it compatible with various materials, including medical devices. However, its usage entails significant risks due to its flammable, toxic, and carcinogenic properties.
Ethylene oxide sterilization hinges on four pivotal factors: gas concentration, humidity, temperature, and duration. By disrupting the DNA of microorganisms, ETO halts their reproduction, ensuring consistent delivery of safe and sterile products to the market.
Process Control and Other Features:
1. AUTOMATIC CONTROL SYSTEM:
The Machine is equipped with automatic control system by Pre-programmed PLC.
Note:- Dual (Automatic & Manual operating) control system will beprovided at extra cost with this model.
2. PRE-PROGAMMED CYCLES:
The sterilizer has equipped 4 pre-programmed cycles as mentioned below:
a) 3 Cartridge Cycles:
Warm Cycle
Cold Cycle
Customer Flexible Cycle
b) 1 Cylinder Cycle.
3. PASSWORD PROTECTION:
a) The sterilizer settings are password protected.
b) The date and time settings are password protected, and User is able to update date and
time settings on the screen for better record keeping.
4. FULLY AUTOMATIC OPERATION NEGATIVE PRESSURE CYCLE (100% ETO CARTRIDGE) CYCLE DESCRIPTION:
a) Vacuuming of chamber to set value
b) Leak hold period
c) Cartridge punching /gas purging depending on the cycle selected
d) Sterilization hold period for set time
e) Exhaust and aeration for no. of aeration pulses set
5. EACH CYCLE FOLLOWS THE BELOW MENTIONED PHASES:
a) CONDITIONING: Chambers are heated by electrical resistance or by steam contained in a jacket surrounding the chamber.
b) INITIAL EVACUATION: To safely deliver the 100% ethylene oxide process, at least 97 percent of the air is removed from the chamber and is achieved byperforming a series of partial vacuums and pulling deep vacuum using vacuum pump.
c) HUMIDIFICATION: This is accomplished by adding humidity in the form of steam injections after the addition of steam, the product is allowed to dwell or soak for time required to replace the moisture lost from the evacuation phase.
d) EXPOSURE (STERILIZING): liquid ethylene oxide is first heated into a gaseous phase, then injected into the chamber.
e) EXHAUST/DEGASSING: After the exposure phase of the process, all gas are removed from the chamber until the levels of EO fall below the flammable limit for the gas (3 percent or 30, 000 ppm).
a) AERATION: The sterilized contents are aerated to remove ETO residue withinthe sterilizer chamber.
6. AUTOMATIC FAULT DETECTION SYSTEM:
a) The Machine is equipped with Faulty Sensor Detection system and user is visually notified of the same.
b) The machine is equipped with temperature detection system and the user will be notified in case the machine fails to reach the set temperature or if the temperature overshoots
c) The machine is equipped with water detection system. The user will be notified in case the water level is not optimum for regular machine operation.
7. PRINTER:
a) The user is record the progress of the sterilization cycle in real time therefore online real time printing facility will be provided.
b) The user can print the following details:
Date and time of the cycle.
Status of the cycle.
Temperature at drain and chamber pressure at every event change and during sterile hold it can be printed at desired interval, minimum of 1 minute.
All undesired situation (faults) are to be printed at the time of fault generation.
8. CYCLE DATA STORAGE FACILITY:
Historical cycle data gets stored in the memory of sterilizer which can be easily retrieved at will through USB.
9. STATE OF THE ART – VACUUM SYSTEM:
Heavy Duty Branded- Oil free diaphragm type vacuum pump is provided to remove the air/gas from the chamber.
10. AUTOMATIC HUMIDIFICATION SYSTEM:
The machine is equipped with inbuilt automatic humidification system and is able to maintain humidity in the chamber within the range of 80 to 85% automatically controlled by wet and dry humidity transducer system.
11. AUTOMATIC LOW WATER CUT OFF SYSTEM:
The machine is equipped with automatic low water cut-off system. When the water level decreases below a threshold then heating elements will automatically turn off. The user will be notified in case the water level is not optimum for regular machine operation.
12. AUTOMATIC PRESSURE CONTROL SYSTEM:
The sterilizer has automatic pressure control switch which automatically maintain the desired pressure and restricts any abnormal/ unsafe rise of pressure in the sterilizer.
13. AIRTIGHT SEALING:
Heavy Duty branded silicon gasket will be provided for airtight seal.
14. PNEUMATIC VALVE:
The sterilizer is equipped with branded pneumatic valves for automatic process.
15. ANALOGUE GAUGES:
The Sterilizer is fitted with branded NABH calibrated analogue chamber pressure (Compound) gauge to measure the pressure and vacuum of chamber.
16. CHAMBER TEMPERATURE CONTROLLER:
The Sterilizer is fitted with chamber temperature controller to control the chambertemperature.
17. HUMIDIFICATION TEMPERATURE CONTROLLER:
The Sterilizer is fitted with humidification temperature controller to control thechamber temperature.
18. CATALYTIC EXHAUST CONVERTER:
Catalytic convertor is provided to treat the ETO gas before exhaust.
19. FLUSH MOUNTING:
The machine is equipped with SS 304 flush mounting only (Front side) Non-Sterile side.
20. EUROPIAN STYLE PANELLING:
The sterilizer has fully covered design with panelling back and both the sides RHS & LHS.
21. DOOR FEATURE:
a) The sterilizer has single door with wing nut locking system.
b) Door opening safety:
Door will be open only at the end of cycle and only when chamber is at room pressure and desired chamber temperature is reached.
22. SAFETY FEATURES:
22.1 EMERGENCY STOP:
An Easily Accessible emergency stop is provided on main control panel door.
22.2 POWER FAILURE:
Equipment settings will not get disturbed due to power failure and will not restart without human intervention.
22.3 FAILURE MODE DETECTION:
Equipment indicate fault through audio / visual signal with details on Control panel/PLC (Failure mode detection).
22.4 HYDRAULIC TESTING:
Complete sterilizer including chamber, jacket, and piping are hydraulically tested at 2 times the working pressure.
22.5 AUTOMATIC DUAL DOOR LOCKING SYSTEM:
5.a) AUTOMATIC ELECTRONIC DOOR LOCK:
The sterilizer is fitted automatic electronic door locking mechanism system.
5.b) AUTOMATIC MECHANICAL DOOR LOCK:
The sterilizer is equipped with automatic mechanical door locking system which prevents the user to open the door until the atmospheric pressure is achieved inside the chamber.
22.6 SAFETY VALVE:
The sterilizer is fitted with pop up type dual safety valves.
22.7 INSULATION:
The sterilizer is completely insulated to prevent any burns caused due to accidental contact to sterilizer body
DESIGN AND FEATURES:
Spray gun rinse unit has designed for connection to water or compressed air, to use for assisted cleaning of pipettes, catheters, cannulas, syringes etc. The spray-gun is include tubing and different tips and nozzles for the various cleaning purposes, e.g.:-
Syringes and cannulas with Record cone.
Measuring and blood pipettes.
Catheters and small pipes.
Drainage tubing.
Syringes and cannulas with Lure cone.
Spray jet for rapid instrument cleaning.
Bottles and Erlenmeyer flasks o water jet pumps for suction cleaning.
All appliances are stored within easy reach on a special wall-mounted rack (included). A special wall-mounted rack a part of standard supply to store all appliances within easy reach.
All tips is get easily locked to the spray gun by a safety cone. The gun grip is heat-insulated. The water/air pressure is released, regulated and fully controlled by the spray-gun trigger (adapted to a 1/2" connection).
Process Control and Other Features:
1. STATE OF THE ART – VACUUM SYSTEM:
Heavy Duty Branded- Oil free diaphragm type vacuum pump is provided to remove the air/gas from the chamber.
2. AUTOMATIC HUMIDIFICATION SYSTEM:
The machine is equipped with inbuilt automatic humidification system and is able to maintain humidity in the chamber within the range of 80 to 85% automatically controlled by wet and dry humidity transducer system.
3. AUTOMATIC LOW WATER CUT OFF SYSTEM:
The machine is equipped with automatic low water cut-off system. When the water level decreases below a threshold then heating elements will automatically turn off. The user will be notified in case the water level is not optimum for regular machine operation.
4. AUTOMATIC PRESSURE CONTROL SYSTEM:
Pressure of the sterilant is controlled automatically through pressure sensor.
5. AIRTIGHT SEALING:
Heavy Duty branded silicon gasket will be provided for airtight seal.
6. ANALOGUE GAUGE:
The Sterilizer is fitted with branded NABH calibrated analogue chamber compound gauge to measure the pressure and vacuum of chamber.
7. DIGITAL CHAMBER TEMPERATURE CONTROLLER:
The Sterilizer is fitted with digital chamber temperature controller and indicator to indicate & control the chamber temperature.
8. DIGITAL HUMIDIFICATION TEMPERATURE CONTROLLER:
The Sterilizer is fitted with digital humidification temperature controller and indicator to indicate & control the chamber temperature.
9. CATALYTIC EXHAUST CONVERTER:
Catalytic convertor is provided to treat the ETO gas before exhaust.
10. DOOR FEATURE:
The sterilizer has single door with wing nut locking system.
Door opening safety:
Door will be open only at the end of cycle and only when chamber is at room pressure and desired chamber temperature is reached.
11. FLUSH MOUNTING:
The sterilizer is equipped with SS 304 flush mounting of front side.
12. EUROPIAN STYLE PANELLING:
The sterilizer has fully covered design with panelling back, top and both the sides RHS & LHS.
13. SAFETY FEATURES:
13.1 HYDRAULIC TESTING:
Complete sterilizer including chamber and piping are hydraulically tested at 2 times the working pressure.
13.2 SAFETY VALVE:
The sterilizer is fitted with spring loaded safety valve.
13.3 INSULATION:
The sterilizer is completely insulated to prevent any burns caused due to accidental contact to sterilizer body.
FEATURES:
Process Control and Other Features:
1. AUTOMATIC LOW WATER CUT OFF SYSTEM:
The Sterilizer is equipped with automatic low water cut-off system. When the water level decreases below a threshold then heating elements will automatically turn off. The user will be notified in case the water level is not optimum for regular machine operation.
2. AUTOMATIC PRESSURE CONTROL SYSTEM:
The sterilizer has automatic pressure control switch which automatically maintain the desired pressure and restricts any abnormal/ unsafe rise of pressure in the sterilizer.
3. WATER LEVEL INDICATOR:
The sterilizer is equipped with water level gauge glass to indicate the water level.
4. ANALOGUE GAUGES:
a) The Sterilizer is fitted with branded NABH calibrated analogue jacket pressure gauge to measure the each boiler pressure.
b) The Sterilizer is fitted with branded NABH calibrated analogue temperature gauge to measure the each boiler pressure.
5. SAFETY FEATURES:
5.1 AUTOMATIC SHUT OFF VALVE:
Prevents hazardous leakage of hot steam and boiled water due to breakage of glass tube.
5.2 HYDRAULIC TESTING:
Complete sterilizer including chamber, jacket, and piping are hydraulically tested at 2 times the working pressure.
5.3 SAFETY VALVE:
The sterilizer is fitted with spring loaded dual safety valves.
Steri has been at the forefront of manufacturing sterilization and infection control products for over 60 years. With wide business base, complete in-house design and manufacturing facility, we have developed a reputation for our quality and reliability. At steri we provide excellent products and exceptional customer service to our clients. Our experience with over 50, 000 installations and strive for excellence has resulted in steri being a pioneer in the field.
Rust-Resistant Construction
High-Quality Stainless Steel is used to avoid corrosion and contamination.
High Purity Water
Water Distillation equipment provides pyrogen free distillate.
Air Tight Sealing
Joint less steam resistant silicon gasket for effective sealing and safety.
Narrow and Compact
The slim design allows the machine to fit in a small space for maximum area utilization.
Green Energy Design
A special system is designed to reuse heated cooling water that considerably reduces the energy consumption resulting in saving electricity.
Steri Eco Aqua-Save Technology
Steri water distillation system is equipped with water-saving technology to save every drop of water being utilized.
Unique Condenser System
The condenser design ensures that the droplets of distilled water remain in contact with the cooling coil for the longest possible time ensuring maximum energy transfer.
Advanced Sensor System
Steri Water Distillation comes equipped with state of the art smart electronic water sensor for accurate detection and heater protection by controlling the water.
Process Control and Other Features:
1. BASKETS:
a) The basket rack is built up as a modular system with storage consoles suitable for storage of sterilizing baskets.
b) The section is suitable for holding 8 baskets.
2. RACK AND FRAME:
a) The entire rack is galvanized and finished smoothly for easy handling and cleaning.
b) The Unit is made of steel profile 50x30 mm welded both sides of a sturdy pillar made of steel profile 50x50 mm.
c) All sides are open.
3. BULLET FEET:
The pillar is mounted on bullet feet suitable for adjusting + 25 mm.
Process Control and Other Features:
1. BASKETS:
a) The basket rack is built up as a modular system with storage consoles suitable for storage of sterilizing baskets.
b) The equipment is dividing into two sections and each section is suitable for holding 8 baskets.
2. RACK AND FRAME:
a) The entire rack is galvanized and finished smoothly for easy handling and cleaning.
b) The Unit is made of steel profile 50x30 mm welded both sides of a sturdy pillar made of steel profile 50x50 mm.
c) All sides are open.
3. BULLET FEET:
The pillar is mounted on bullet feet suitable for adjusting + 25 mm.
Process Control and Other Features:
1. MANUAL KNEW OPERATED SYSTEM:
Flow of water is controlled by Knew operated switch.
2. FOOT OPERATED SWITCH:
Flow & quantity of liquid soap / disinfectant is controlled with the help of foot operated switch.
3. GEYSER SYSTEM:
The machine is equipped with the geyser system.
4. WATER PUMP:
The machine is fitted with the water pump.
5. AIR GUN SYSTEM:
Scrub station is fitted with the air gun system for drying.
6. EUROPEAN STYLE FULLY COVERED:
Scrub station is fully covered design with panelling Front, back and both the sides RHS & LHS.
7. WATER TANK WITH ACCESSORIES:
The unit also consists of a SS 304 water tank, water heating system and water pump along with manual valves for hot and cold mixing.
8. SPECIAL DESIGNED:
The complete unit is made of SS 304 (16 gauge) having rounded corners is provided for easy cleaning.
FEATURES:
The trolley contains 2 shelves.
The frame is mounted on 4 nos. heavy duty swivelling castors for easy movement in narrow passages.
Stainless steel table trolley has 2 shelves for use in the sterile goods store when batching sterile goods for delivery and also for short internal transport.
The trolley is also suitable for use as a temporary storage or utility space at packing tables.
Process Control and Other Features:
1. AUTOMATIC LOW WATER CUT OFF SYSTEM:
The Sterilizer is equipped with automatic low water cut-off system. When the water level decreases below a threshold then heating elements will automatically turn off. The user will be notified in case the water level is not optimum for regular machine operation.
2. AUTOMATIC PRESSURE CONTROL SYSTEM:
The sterilizer has automatic pressure control switch which automatically maintain the desired pressure and restricts any abnormal/ unsafe rise of pressure in the sterilizer.
3. DOOR FEATURE:
a) The sterilizer has SS 304 door with remote operated system.
b) The sterilizer door is equipped with hydraulic lid which closes the sterilizer smoothly.
4. THERMOSTATICALLY CONTROLLED:
The sterilizer is controlled by thermostatically.
5. INSTRUMENTS TRAY WITH HANGER SYSTEM:
The sterilizer is equipped with the instruments tray and a hanger system to dip the instrument full inside water.
Process Control and Other Features:
1. TIMER CONTROLLER:
Drying timer controlled drying cabinet with single door.
The system has drying time range of 0-90 minutes.
2. ADJUSTABLE TEMPERATURE SYSTEM:
The system has adjustable adjustment temperature range of 60-90°C.
3. POWER CONSUMPTION:
Suitable for electrical operation on 230 volts, single phase, 15 Amps. Having Electric load of 3.6 KW.