We offer a complete product range of Bod Meter, TURBIDITY METER, PH Meter, Digital PH Meter and Dissolved Oxygen Meter
Description
Analyzes of COD, TOC, total phosphate and total nitrogen require chemical mineralization (or digestion) withheating. The RD010 heating reactor offers programs at several temperatures (100/120/150 C) and reactiontimes (30, 60 and 120 minutes) as well as continuous operation. It has 24 stations for 16 mm diameter tubes.
It operates on 220 V and 110 V (voltage switch on the back of the device)
Since safety is highly important, the heating reactor has a protective cover and is equipped with precise
temperature control and overheating protection. An alarm sounds when the digestion is complete and thereactor switches off automatically.
For the correct handling of the tubes after heating, it is recommended to use a clamp adapted to catch themand to deposit them in a stand for cooling.