We offer a complete product range of TE2 Temperature Transmitter, Pressure Transmitter, Calorimetric Flow Transmitter, TER8 Temperature Transmitter and PL20H Level Switch
Process temperatureRefer to section "Operating conditions"
Pt100 accuracy class (EN 60751)B (± 0.3 °C at 0 °C), ± (0.3 + 0.005 × t)°C, A (± 0.15 °C at 0 °C), ± (0.15 + 0.002 × t)°C, 1/3 B (± 0.1 °C at 0 °C), ± 1/3 × (0.3 + 0.005 × t)°C, 1/6 B (± 0.05 °C at 0 °C), ± 1/6 × (0.3 + 0.005 × t)°C,
Thermal response time, T50RTD element only- ≤ 1.0 s , conical sensor, ≤ 1.3 s , Ø3 mm, ≤ 2.5 s , Ø4 mm, ≤ 3.0 s , Ø6 mm,
Process pressureRefer to section "Operating conditions"
Sensor length≤ 3000 mm
Sensor tube materialAISI 316L (1.4404)
Surface roughness wetted partsRa ≤ 0.8 µm
Degree of protection (EN 60529)IP 65 , with DIN EN 175301-803 A (DIN 43650 A), 4-pin IP 67 , with M12-A, 4-pin
Humidity≤ 100 % RH , condensing
Long-term test (transmitter)IEC 770 6.3.2
Vibration (sinusoidal) (EN 60068-2-6)1.6 mm p-p (2 … 25 Hz), 4 g (25 … 100 Hz), 1 octave / min.
Without transmitter1 x Pt100, 4-wire
With transmitter4 … 20 mA , 2-wire
EMCEN 61000-6-2, EN 61000-6-3, EN 61326-1, NE21
HygieneFDA (21 CFR 177.2415) 3-A (74-07)
Railway applicationsEN 50155
Explosion protectionEx ia Simple apparatus, gas and dust
Process temperature115 … 135 °C , < 1 h -40 … 115 °C , permanent
Output range-50 … 150 °C , programmable
Min. output span25 °C
Pt100 accuracy class (EN 60751)B (± 0.3 °C at 0 °C), ± (0.3 + 0.005 × t)°C, A (± 0.15 °C at 0 °C), ± (0.15 + 0.002 × t)°C, 1/3 B (± 0.1 °C at 0 °C), ± 1/3 × (0.3 + 0.005 × t)°C, 1/6 B (± 0.05 °C at 0 °C), ± 1/6 × (0.3 + 0.005 × t)°C
Standard error of measurement (BFSL) ± 0.04 % FSR , ± 0.1 % FSR ± 0.2 % FSR Including non-linearity, hysteresis and non-repeatability according BFSL For turndown, multiply this value by the applied turndown ratio
Max. measuring error ± 0.1 % FSR, , ± 0.25 % FSR ± 0.5 % FSR Including zero-point and span error, nonlinearity (by terminal base line), hysteresis and non-repeatability (EN 61298-2) For turndown, multiply this value by the applied turndown ratio
Temperature coefficient ≤ 0.03 % FSR/10 K , measuring span ≤ 0.03 % FSR/10 K , zero point
Compensated temperature range -40 … 85 °C
Long term stability ≤ 0.1 % FSR/a , measuring range > 1 bar ≤ 1 mbar , measuring range ≤ 1 bar
Standard error of measurement (BFSL) ± 0.04 % FSR ± 0.1 % FSR Including non-linearity, hysteresis and non-repeatability according BFSL
Max. measuring error ± 0.1 % FSR ± 0.25 % FSR Including zero-point and span error, nonlinearity (by terminal base line), hysteresis and non-repeatability (EN 61298-2) For turndown, multiply this value by the applied turndown ratio
Temperature coefficient ≤ 0.03 % FSR/10 K , measuring span ≤ 0.03 % FSR/10 K , zero point
Compensated temperature range -40 … 85 °C
Long term stability ≤ 0.1 % FSR/a , measuring range > 1 bar ≤ 1 mbar , measuring range ≤ 1 bar
Max. turndown ratio 05:01:00 AM
Rise time (10 ... 90 %) ≤ 5 ms
IO-Link version 1.1
IO-Link port type Class A
Baud rate 38, 4 kbaud (COM2)
Cycle time ≥ 2.3 ms
Process data length 24 bit
SIO-mode No
Process data (cyclic) Process pressure
Diagnostic data (acyclic) Process temperature
Explosion protection ATEX II 1/2G Ex ia IIC T4/T6 Ga/Gb ATEX II 1D Ex ia IIIC T107 °C IP6X Da ATEX II 1G Ex ia IIC T4/T6 Ga
Standard error of measurement (BFSL)0.04 % FSR, 0.1 % FSR, 0.2 % FSR, Including non-linearity, hysteresis and non-repeatability according BFSL For turndown, multiply this value by the applied turndown ratio
Max. measuring error± 0.1 % FSR, ± 0.25 % FSR, ± 0.5 % FSR, Including zero-point and span error, nonlinearity (by terminal base line), hysteresis and non-repeatability (EN 61298-2) For turndown, multiply this value by the applied turndown ratio
Temperature coefficient≤ 0.03 % FSR/10 K , measuring span, ≤ 0.03 % FSR/10 K , zero point
Compensated temperature range-40 … 85 °C
Long term stability≤ 0.1 % FSR/a
Max. turndown ratio05:01:00 AM
Rise time (10 ... 90 %)≤ 5 ms
Explosion protectionATEX II 1/2G Ex ia IIC T4/T6 Ga/Gb, ATEX II 1D Ex ia IIIC T107 °C IP6X Da, ATEX II 1G Ex ia IIC T4/T6 Ga
Process temperatureRefer to section "Operating conditions"
Pt100 accuracy class (EN 60751)B (± 0.3 °C at 0 °C), ± (0.3 + 0.005 × t)°C, A (± 0.15 °C at 0 °C), ± (0.15 + 0.002 × t)°C, 1/3 B (± 0.1 °C at 0 °C), ± 1/3 × (0.3 + 0.005 × t)°C, 1/6 B (± 0.05 °C at 0 °C), ± 1/6 × (0.3 + 0.005 × t)°C
Thermal response time, T50< 1.5 s , Ø4 mm, < 6.1 s , Ø6 mm, ≤ < 7.6 s , Ø8 mm
Process pressureRefer to section "Operating conditions"
Sensor length20 … 3000 mm
Sensor tube materialAISI 316L (1.4404)
Surface roughness wetted partsRa ≤ 0.8 µm
Degree of protection (EN 60529)IP 67 IP 69K , with appropriate cable
Humidity< 98 % RH , condensing
Vibration (sinusoidal) (EN 60068-2-6)1.6 mm p-p (2 … 25 Hz), 4 g (25 … 100 Hz), 1 octave / min.
Without transmitter1 x Pt100, 4-wire, 1 x Pt100, 4-wire, 2 x Pt100, 2-wire
With transmitter4 … 20 mA , 2-wire, 4 … 20 mA , 2-wire + HART®
EMCEN 61000-6-2, EN 61000-6-3, EN 61326-1
Hygiene3-A (74-07)
Explosion protectionATEX II 1G Ex ia IIC T4/T5, ATEX II 3G Ex nA IIC T4/T5, Ex ia Simple apparatus, gas and dust
Temperature compensation0.0 … 5.0 % FSR/K , adjustable
Step response time≤ 2.0 s
Sample time≤ 0.4 s
Concentration4 factory set media
Customer defined mediaCustomer defined (30 point lookup table)
TemperatureFree programmable range
Measuring range-20 … 150 °C
Thermal response time, T90≤ 15 s
Max. measuring error± 0.4 K
Process temperature-20 … 140 °C , permanent 140 … 150 °C , max. t < 1 h
Process pressure≤ 25 bar
Connection variantsG 1 A hygienic
Wetted parts materialPEEK Natura
Surface roughness wetted partsRa ≤ 0.8 µm
PropertiesUniversal commands Common-practice commands Conductivity device family commands Device Specific Commands For more information please see ‚HART Field Device Specification‘
ProtocolHCF standard, Rev.7
IO-Link version1.1
Device profileSmart Sensor Profile
Baud rate38, 4 kbaud (COM2)
Cycle time≥ 8, 4 ms
Process data length128 bit
SIO-modeYes
Process data (cyclic)Switch state Signal analog output 1 Signal analog output 2 Temperature Unit temperature Conductivity Concentration Actual measuring range
Degree of protection (EN 60529)IP 67 IP 69K , with appropriate cable
Cable lengths10.0 m 5.0 m 2.5 m
MaterialPUR
Temperature-40 … 80 °C
EMCEN 61326-1
Hygiene3-A (74-07) EHEDG EL Class I FDA (21 CFR 177.2415)
Temperature compensation0.0 … 5.0 % FSR/K , adjustable
Step response time≤ 2.0 s
Sample time≤ 0.4 s
Concentration4 factory set media
Customer defined mediaCustomer defined (30 point lookup table)
TemperatureFree programmable range
Measuring range-20 … 150 °C
Thermal response time, T90≤ 15 s
Max. measuring error± 0.4 K
Process temperature-20 … 140 °C , permanent 140 … 150 °C , max. t < 1 h
Process pressure≤ 25 bar
Connection variantsG 1 A hygienic
Wetted parts materialPEEK Natura
Surface roughness wetted partsRa ≤ 0.8 µm
PropertiesUniversal commands Common-practice commands Conductivity device family commands Device Specific Commands For more information please see ‚HART Field Device Specification‘
ProtocolHCF standard, Rev.7
IO-Link version1.1
Device profileSmart Sensor Profile
Baud rate38, 4 kbaud (COM2)
Cycle time≥ 8, 4 ms
Process data length128 bit
SIO-modeYes
Process data (cyclic)Switch state Signal analog output 1 Signal analog output 2 Temperature Unit temperature Conductivity Concentration Actual measuring range
Degree of protection (EN 60529)IP 67 IP 69K , with appropriate cable
EMCEN 61326-1
Hygiene3-A (74-07) EHEDG EL Class I FDA (21 CFR 177.2415)
Standard error of measurement (BFSL)± 0.04 % FSR , up to 2:1 turndown ratio, ± 0.1 % FSR , up to 4:1 turndown ratio, Including non-linearity, hysteresis and non-repeatability according BFSL For turndown, multiply this value by the applied turndown ratio
Max. measuring error± 0.1 % FSR , up to 2:1 turndown ratio, ± 0.25 % FSR , up to 4:1 turndown ratio Including zero-point and span error, nonlinearity (by terminal base line), hysteresis and non-repeatability (EN 61298-2) For turndown, multiply this value by the ap
Temperature coefficient≤ 0.05 % FSR/10 K , measuring span ≤ 0.05 % FSR/10 K , zero point
Compensated temperature range-40 … 85 °C
Long term stability≤ 0.1 % FSR/a , IEC 770 6.3.2
Max. turndown ratio10:01:00 AM
Rise time (10 ... 90 %)≤ 0.3 s
Sample time≤ 0.3 s
Power-up time< 10 s
Explosion protectionATEX II 1D Ex ia IIIC T100 °C Da, ATEX II 1G Ex ia IIC T5, ATEX II 3G Ex nA IIC T5
Standard error of measurement (BFSL) ± 0.3 % FSR , available for measuring spans > 1 bar ± 0.4 % FSR Including non-linearity, hysteresis and non-repeatability according BFSL
Max. measuring error ± 0.5 % FSR , available for measuring spans > 1 bar ± 1 % FSR Including zero-point and span error, nonlinearity (by terminal base line), hysteresis and non-repeatability (EN 61298-2)
Temperature coefficient ≤ 0.08 % FSR/10 K , measuring span ≤ 0.08 % FSR/10 K , zero point