Company Information
Ask for more detail from the seller
Contact SupplierTEK1-pSQT Pure Steam Quality Test Kit is equipped with all the components that are necessary to perform the required test and report the results in accordance with EN 2855, HTM 20102 and ISO 111343 for the following parameters:
Non-Condensable Gas Value Steam Superheat Value Steam Dryness Value
Non-Condensable Gas Value
Non-condensable gas test is designed to demonstrate that the attainment of sterilization conditions in all parts of a sterilizer load (particularly for porous load items) is not impaired by the presence of non-condensable gases. There are two sight glasses provided for visualization of steam condensate and non-condensable gas level. These sight glasses are supported by zero adjustable, calibrated scales in the background. When the sampled steam is condensed, the non-condensable gases present are released. The Gas component is separated from the cooled condensate. The gas and steam condensate volumes are measured in the sight glasses.
The non-condensable gas value is considered to be acceptable for sterilization purposes on pharmaceutical and clinical sterilizers if the percentage of gas to condensate is less than 3.5%.