The company presents family of radiant warmers to meet the diverse thermo regulatory requirements of patient at your hospital. V-800 Open Care System is designed according to the most recent....
The company presents family of radiant warmers to meet the diverse thermo regulatory requirements of patient at your hospital. V-800 Open Care System is designed according to the most recent international standards and provides the neonate with an adequate and secure environment of its care. This compact and mobile Open Care System with integrated phototherapy system is a must for every well-run and busy hospital. The integrated halogen based Phototherapy system ensures that optimum therapeutic output is available on demand to manage the hyperbilirubinaemia without disturbing the infant.
Features
The controller is based on Microcomputerized system.
Large bright displays are easy to monitor and for viewing from a distance.
Double digital display system provided to measure the Infant temperature and to set the required temperature.
Feather touch keys provided for easy operation.
Bar graph LED system provided to indicate the heater output.
Manual control system provided.
Fault Indications with alarm for Baby probe failure, 1oC temperature high, 1 oC temperature low, High temperature cut-off, Power failure
The Electronic Controller
The controller is based on Microcomputerized system
Large bright displays are easy to monitor and for viewing from a distance
Double digital display system provided to measure the Infant temperature and to set the required temperature
Feather touch keys provided for easy operation
Bar graph LED system provided to indicate the heater output
Manual control system provided
Fault Indications with alarm for
Baby probe failure
1oC temperature high
1oC temperature low
High temperature cut-off
Power failure
Baby Probe
Baby probe is made of highly sensitive sensor and is interchangeable, field calibration taken care of by software. The probe is made up of silicon sheathed cable, which allows easy cleaning
Baby Tray
The large baby tray is made of stainless steel frame with three side acrylic collapsible walls, trendelenburg & fowler system provided
Observation Lamp
A halogen based observation light focus on any area on the bed
Heater
Single element CE approved ceramic heater placed in a parabolic reflector
Other Facilities
Provision of Oxygen cylinder
I V stand provided
Monitor tray provided
Physical ParametersMattress
Length :700 mm
Width : 500 mm
Thickness : 50 mm
Working level from ground : 1020 mm
Baby tray is tilt angle on either side : +/-8 max
Weight Approx : :40 kg
Castors : 100 mm with 2 brake wheels and 2 non - brake wheels
Recommended Environmental Specifications
Operating temperature range : 20oC to 33oC
Storage temperature range : -25oC to 60oC
Operating humidity range : 0 - 100% RH
Skin Temperature Probe
Range : 28oC to 38oC
Accuracy : +/-0.2oC within range
Resolution : 1oC
Probe interchangeability : +/-0.2oC (Servo)
Heater output controllable from 0 to 100% in manual mode (continuously variable)
Phototherapy Unit
Supply to each lamp : 12V, 50W
Irradiance : 6 - 8 µW /cm 2/nm at bed level
Observation Light
Halogen based Power required : 12V, 50 W
Electrical Specifications
Electrical Supply : 230V ± 10%, 50 Hz
Power consumption with Heater at max. output : 750W
Power Consumption
With heater, Phototherapy & observation lamps ON : 1250 W
Electrical Safety
As per IEC 60601-1-2 for medical equipments
Note : Best results obtained if air convective current is low