Contact SupplierHeat plus resq overview:
parts: microphone, status button, pads port,shock button, internal speaker, power button for adults, power button for infants
defibrillator series: nt-381.c nt-381.b
set configuration: defibrillator, battery pack, sealed electrode, pads, heart+resq user guide, protective carrying case, "nt-mpr" cd-rom, bluetooth, adaptor (optional)
waveform: truncated exponential biphasic
energy:150j (adult, impedance 50 omega), 50j (pediatric, impedance 50 omega)
shock charge time:9 seconds (±10%)
shock method:manual button (flashes when ready)
mode conversion: adultchild toggle button, (child mode delivers 50j pediatric shock for patients weighing less than 25kg,
approximately 12 months to 8 years old)
physical: dimensions 8.0cm(h) x 22.0cm(w) x 28.2cm(d), weight 2kg (including battery and pads)
environmental:
water resistance ip55 per iec 60529
temperature operating: 0c to 40c;
storage: -20c to 60c
humidity(operatingstorage): 10% to 95% relative humidity, non-condensing
altitude (operatingstorage): 0 to 5000 meters
drop test:1.5 meters per iec 60068-2-32
ecg:
ecg recording: lead ii
analysis overview: patient evaluated to determine whether ecg rhythm represents shockable or non-shockable state. Shockable rhythms are ventricular fibrillation (vf) and ventricular tachycardia (vt).
Analysis time: approx. 7 seconds (read: about 5 seconds, analyze: within 2 seconds)
data:
stored contents: ecg recordings, operation information (whether shock delivered or not), ambient noise, self-diagnostics history.
Memory: ecg records 5 sec during analysis only (up to 1000 instances), ambient noise: 60 minutes,4 inst of 15 min