MODEL MAMMOMAT 3000 NOVA

Listing ID #3883778

  • POWER NEEDED208 VAC, single phase
  • ROTATION, ° (ILLUMINATION LEVEL, maximum ft-candles)+135, -180, motorized rotation
  • mAs RANGE (X-RAY TUBE)2-600, Mo tube. The mAs range is 2 with the tube.
  • Focal spot size, mm (X-RAY TUBE)0.1 and 0.3
  • Supply Type Manufacturer, Exporter, Supplier
  • Preferred Buyer Location All over the world
View More Details
Send Enquiry

Company Information

  • Member Since 13 Years
  • Nature of Business Retailer

Ask for more detail from the seller

Contact Supplier

Product Details no_img_icon

  • AEC DETECTOR (POSITIONING ASSEMBLY) Solid-state, sandwich design
  • L x w, cm (in) (CONTROL CONSOLE) 195 x 65 (76.8 x 25.6)
  • Bucky (GENERAL RADIOGRAPHY) optional 18 x 24 cm, 24 x 30 cm; mounted on pivoting arms
  • Focal spot size, mm (X-RAY TUBE) 0.1 and 0.3
  • DIMENSIONS (HXWXD) CM, (IN) (DISPLAY) 217 x 52 x 22.5 (85.4 x 20.5 x 9) stand, 196 x 70 x 34 (77 x 27.6 x 13.4) generator
  • Anode type (X-RAY TUBE) Rotating 8,800 rpm
  • mAs RANGE (X-RAY TUBE) 2-600, Mo tube. The mAs range is 2 with the tube.
  • GENERATOR High-frequency, single- or two-phase power supply
  • RADIATION SHIELD (X-RAY TUBE) Yes
  • POWER NEEDED 208 VAC, single phase
  • ROTATION, ° (ILLUMINATION LEVEL, maximum ft-candles) +135, -180, motorized rotation
  • Kv range (X-RAY GENERATOR) 23increments of 1 kV
  • Vertical axis, cm (in) (POSITIONING ASSEMBLY) 65-135 (25.6-53.1) motorized

  • Dual Bucky System: Small 18 x 24 and Large 24 x 30
  • Generator: High Frequency X-Ray Generator
  • Tube: Dual Track Molybdenum/Rhodium X-Ray
  • Focal Spots: .01mm and .3mm Focal Spot Sizes
  • Exposure Modes: Three Exposure Modes (Automatic
  • Optimization of Parameters (AOP), Fully Automatic, and AECSemi Automatic/Manual
  • Magnification: 1.3 and 1.9 Magnification Factors – Single Platform
  • Telescopic Motorized Positioning Column: Command Buttons
  • Located on Each Side and Top of the Column for Easy Access
  • Digital Display: Compression Pressure, Breast Thickness, Size
  • of X-Ray Field, and Magnification Factor
  • Foot Pedals: Motorized Compression Controlled by Two
  • Compression/Decompression Foot Pedals with Manual Fine Tunning
  • Compression Paddles: Slide Mount for Stability/Easy of Mounting
  • System Also Includes: Image Receptor, Control Console,
  • Lead/Glass Operator Radiation Shield, Face Guard, Operator
  • Keyboard/Console, DataFlash Patient ID System, Operator
Generator
  • High-voltage wave form: Multipulse
  • Exposure voltage : 23kV to 35kV, adjustable in increments of 1kV
  • mAs range : Mo X-ray tube: 2mAs to 560 mAs (mAs-mode)up to 600mAs (AEC-mode)
  • Mo/W X-ray tube : 2mAs to 710 mAs (mAs-mode) up to 752 mAs (AEC-mode)
  • Exposure times : From 10 ms to 4s , From 10 ms to 7s for magnification techniqueAutomatic exposure control
  • Automatic exposure control (AEC): Microprocessor controlled, transparency compensation
  • Detectors: Semi conductor detector
  • Accuracy of optical density (2-6 cm): ± 0.15 OD from the mean optical density for predefined film screen combinations (separately for each grid table, magnification table and anode-filter combination at appropriate clinical kV).
  • Density correction : ± 3 exposure points, adjustable in 1/8 step increments
  • Film-screen combinations : 2; User selectable on the generator control panel (H/D).


Interested in this product? Ask for more details & Latest Price from seller
Send Enquiry

Company Details close-icon

To be the most focused high technology solutions provider, based on enduring customer success, creating value in society through an inspired team of professionals, committed to good governance and finest operating standards.
Suntion Team of Professions Include the expertise of :
Vision
The Vision of the Company is to become the leader in the high technology industry; respected for the value created in Society, be driven by our Corporate Social Responsibility to establish and enhance industry standards through innovation, integrity and professionalism.
Company Values
The Company value system requires profits to be generated only through “righteous” products, service and work environment and that all stake holders benefit and experience “win-win” from the value addition process.
Quality Policy
We are committed to Providing Quality Products and Services that ensure Customer Success.
We strive to ensure that the Customer Requirements we obtain are Correct, Current Clear and Complete.
We are committed to maintain and continually improve our systems and processes to enhance our reputation with our customers.
We will strive continuously to improve our service quality by the involvement and accountability of a team of empowered well trained employees.
Dedication to Quality as an Organization wide responsibility is a way of life at Analytical Instruments Group
  • Nature of Business Manufacturer / Exporter / Supplier / Retailer
Tell us your Buy Requirement to Get Instant Response
Tell us what you need?

By clicking Get Quotes Now, I accept the T&C and Privacy Policy.

Looking for MODEL MAMMOMAT 3000 NOVA?

Quantity
Seller Contact Details

Find Seller from near by Cities

Waiting for permission
To search by voice, go to your browser settings and allow access to microphone

Allow microphone access to search with voice