product description :
our durable lp series of portble volume ventilator have a repurarion for dependability. They are built with a patented brushes motor,
features
front panel controls :
mode: standby, assist/control, simv or pressure cycle
low pressure alarm: set limit for low pressure alarm 2 to 32 cmh2o in increments of 2 cmh2o
high pressure alarm limit: (assist/control, simv modes) or limit
for pressure cycle mode: 15 to 90 cmh2o in increments of 5 cmh2o
volume: push-to-turn knob: 100 to 2200 ml
breath rate: 1 to 20 bpm in increments of 1 bpm and 22 to 38
bpm in increments of 2 bpm
inspiratory time: 0.5 to1.0 sec in increments of 0.1 sec; 1.2 sec; and 1.5 to 5.5 sec in increments of 0.5 sec
breathing effort: push-to-turn knob sets pressure level to detect breathing effort (-10 cmh2o/hpa to +10 cmh2o)
front panel input
patient pressure: port for connection to proximal pressure line of patient circuit
front panel input
patient pressure: port for connection to proximal pressure line of patient circuit
front panel output
patient air: 22 mm x 15 mm tube for connection to bacteria filter
exhalation valve: port for connection to exhalation valve of patient circuit
indicators, normal events
patient pressure meter: displays patient pressure -10 to +100 cmh2o; can also display battery charge and machine hrs of operation
breathing effort: green led activated by adequate patient breathing effort
power: leds indicate operating power source: green ac pwr/batt
chg, amber external battery, flashing amber internal battery
alarms: flashing red leds: low pressure/apnea, low power, high pressure, setting error, power switchover
audible alarms
pulsating tone: low pressure/apnea; low power, high pressure,
setting error, power switchover
steady tone: loss of microprocessor control
reminder tone: every 5 mins when powered by internal battery,
each time accessory printer generates report
rear panel controls
ac circuit breaker: internal to ac power switch
(1 for on; 0 for off)
rear panel inputs
inlet filter: intake for patient air; screw off cap for filter change
external battery: connection for 12
vdc battery communication port: calibration information during service procedure