Contact SupplierCIL provides 'Ingress Protection Testing' services in Coimbatore. IP65 Testing, IP67 Testing, IP55 Testing, etc.
Ingress Protection Laboratory
Ingress protection test lab is equipped with dust test chamber to perform IP 5X and 6X to accommodate the test panels/ samples of size up to 2.2 m x 1.8 m x 2.5 m (W x L x H) and water test facility to perform tests up to X8 test. The water immersion tank is available to conduct IP X7 and X8 for sample of size up to 1.2m x 1.2m x 1.6m (W x L x H).
The Lab under takes the test to protect against solid foreign object, hazardous objects and liquid (water) and having test facility ranging from IP 11 to IP 68.
AUGMENTED TEST FACILITY
Ingress protection laboratory is being augmented with state of art test facility where in a huge dust chamber being commissioned to accommodate the large panels / samples of size up to 5m x 4.5m x 4.5m (W x L x H) and of weight up to 15 Tons. The lab is equipped with 15 Tons EOT crane for hassle free material handling. The laboratory will accommodate large test sample like Heavy motor, Compact substations, electrical panels etc. The laboratory will also have water tank to conduct X7 and X8 immersion tests for the samples up to 3.5m x 3.5m x 4m (W x L x H).
To meet the customers, demand and to fully comply IEC 60529 standard, IP X9 test facility is being established for first of its kind. IP X9 test is performed in water having high temperature and pressure.
Establishment of Additional Test facility (IPX9)
The CDG established Degree of Protection IP Test equipment for second characteristics numeral IP X9 as per IEC 60529 -2013. IP X9 test is performed using water having high temperature and pressure. The Test set up for IPX9
The Technical specifications of the Test set up are furnished below.
Dimensions of the Chamber: 1200mm(L) X 1200 MM(W)X1200MM(H)
Temperature Range: 80 ± 5ºC
Power Supply: 230V AC, Single Phase, 50 HZ
There are four nozzles in the Test chamber with spray angles at 0º, 30°, 60º, 90°. The sequential high temp water spray will be injected for 30 sec by each nozzle while the test sample is kept over turn table of 5rpm±1 rpm. The flow rate of water is 15 liters ± 1litre /minute
Ingress protection testing tests the level of protection an enclosure provides against intrusions. One such set of ingress protection tests involves exposing the equipment under test, to high-pressure and high-temperature water jets. These tests are numbered IPX9. The first digit, represented by the “X”, ranges from 0 to 6 and specifies the protection from solid particles. The second digit, in the case of the high-pressure and high temperature water jets ingress protection test, would always be at X 9.
The CDG in general carry out testing and certification for IP rating IP69
The CDG carry out all the possible IP X9 ingress protection code combination.
The CDG/EATD Engineers having experience in helping manufacturers to determine ingress points in their products and also identifying required solutions
We also have significant expertise in interpreting the standards, which often provides manufacturers with flexibility in determining acceptable amounts of ingress. Just because some liquid may enter an enclosure, does not automatically mean that the unit has not met the requirements