Contact SupplierInput | Two wire current signal (μA) |
---|---|
Transducer Compatibility | CHS-HTR-PH-100-2W, Mounted within 1 mt. distance. |
Max Negative offset I/P current | ~ 20% of Span set. |
Signal amp I/P current | Max +/-10 pA |
Sensor Input Isolation Power | Galvanic isolation from the Two wire loop 10Mohm at 500 Vdc test voltage |
Electrical Non- Linearity | 1% of F.S. |
Offset & Span drift | 100 ppm of reading /°C |
Minimum operating Voltage | 12 Vdc |
Absolute Max. Operating Voltage | 36 Vdc |
Reverse polarity protection | Reverse Polarity Protected |
Span & Zero Settings | 4 mA and 20 mA currents can be set by Presets R13 & R15 resp. |
Termination used | Signal input on a Teflon Block, Cable shield as a sensor ground. |
Calibration | By 4 position DIP switch +ENTER Key |
Operating Temp. | 0oC to +55o C |
Humidity | 98%RH at40 deg C Non-condensing |
Size | ~ 90 mm ( Dia) X 35 mm ( H ) |