Listing ID #918353
Company Information
Ask for more detail from the seller
Contact SupplierWhy is Calibration necessary?
Temperature can not be measured entirely without error, so it is necessary to quantify these errors by calibration of the probe, with or without its associated display against the reference standard in the respective temperature range. Calibration is essential where traceability to a National Standard or International Standard in compliance to industrial requirement in ISO-9000 specification.
Labin Temperature Heating/Water/Cooling/Clinical Baths may be used for temperature application like Calibration Measurement and Comparison of the Thermometers and Associate Instruments in the given range -50 C to +300 C. An electronic proportional band control adapts the heat supplied to the bath for thermal stability and uniformity requirements of the Bath for Comparison/Calibration.
Specifications |
PRS 300-H |
PRS 100-H |
PRS 20-H |
PRS 50-H Clinical Thermometer |
---|---|---|---|---|
Working |
50 to 300oC |
Amb. +5 to 100oC |
-20 to Amb. oC |
30 to 50oC |
Resolution of Display |
±0.1oC |
±0.01oC |
±0.1oC |
±0.01oC |
Overall Temp. Stability |
±0.05oC |
±0.02oC |
±0.1oC |
±0.02oC |
Uniformity |
±0.05oC |
±0.02oC |
±0.1oC |
±0.02oC |
Temperature Control |
Microprocessor |
Microprocessor |
Microprocessor |
Microprocessor |
Heater Wattage |
1000 Watt |
750 Watt |
750 Watt |
250 Watt |
PT-100 Sensor |
3-Leads |
3-Leads |
3-Leads |
3-Leads |
Safety Installation |
Drain Plug & Overflow Outlet for Expanded Bath Liquid |
Drain Plug & Overflow Outlet for Expanded Bath Liquid |
Drain Plug & Overflow Outlet for Expanded Bath Liquid |
Drain Plug & Overflow Outlet for Expanded Bath Liquid |
Stirrer Motor Power |
1/8 hp, 1425 rpm |
1/8 hp, 1425 rpm |
1/8 hp, 1425 rpm |
1/8 hp, 1425 rpm |
Bath Opening Dia x Depth |
100 x 500 mm |
100 x 500 mm |
150 x 500 mm |
160 x 200 mm |
Overall Dimension |
400 x 400 x 600 mm |
400 x 400 x 600 mm |
450 x 450 x 600 mm |
300 x 300 x 300 mm |
Filling Volume |
4.5 to 5.5 ltr |
4.5 to 5.5 ltr |
9 to 10 ltr |
4.0 ltr. |
Insulation - Fiber Wool & Puff |
Not Less Than 135 mm |
Not Less Than 135 mm |
Not Less Than 135 mm |
Not Less Than 100 mm |
Mains Power Connections |
220-240/50 V/Hz |
220-240/50 V/Hz |
220-240/50 V/Hz |
220-240/50 V/Hz |
Power Consumption |
1500 Watt |
850 Watt |
3000 Watt |
500 Watt |
Permissible Ambient Temp. |
10 to 35oC |
10 to 35oC |
10 to 35oC |
10 to 27oC |
Bath Material |
S. S. 304 Grade / Powder Coating |
---|---|
Wall Thickness of Sheet Material |
1 mm. |
Insulation |
Fiber Wool & Puff. |
Thickness of Insulation |
Not less than 135 mm. |
Stirrer Type |
Screw Type |
Stirrer |
Brass Stirrer L 400-425 mm x Ø 50 mm |
Cover Plate of Bath Chamber |
Hylem Sheet 19 mm thick to hold 5-6 Thermometers |
Ø of Holes in the Cover Plate |
mm 8(2), 6(2), 10(1), 12(1) other than for Control RTD and Heating Rod. |
Thermometers Holding Clips |
Phosphor Bronze |
Stirrer Motor |
¼ h.p., 1425 R.P.M., Crompton Make. |
Heating System |
Nichrom Heater Winding 1200 W for Low Temp. Bath and 1000 W for Water & Oil Baths with 12-5A/230V AC. |
Temp. Controller |
Digital PID Temp. Indicator-cum-Controller & LABIN Make Digital Temp Indicator with Display 0.01oC upto 199oC. |
Temperature Sensor |
RTD Pt-100, Probe Type |
Level Control of Fluid |
Exit Pipe will be provided for Overflow of Bath Liquid. |
Frame/Stand |
A Stable Wheeled Metallic Frame to hold the Bath alongwith Control Panel will be provided. Proper arrangement for Tube Light will also be provided. |
Compressor |
Sriram/Kirloskar Make. |