    Gas Meter Calibration Test Bench

    • Supply TypeManufacturer, Exporter, Supplier, Retailer
    • Preferred Buyer Location All over the world

    The ‘Calibration Standard’ of the Test Bench is “Bell Prover”, which is considered a Primary (Calibration) Standard to Test & Calibrate Domestic Gas Meters (& Gas Meters used in....
    View More Details
    Send Enquiry

    Company Information

    • calendar Member Since 11 Years
    • building Nature of Business Retailer

    Ask for more detail from the seller

    Contact Supplier
    Report incorrect details

    Product Details no_img_icon

    The ‘Calibration Standard’ of the Test Bench is “Bell Prover”, which is considered a Primary (Calibration) Standard to Test & Calibrate Domestic Gas Meters (& Gas Meters used in Hotels/Restaurants etc.). The device is ideal for OEMs, Calibration Laboratories & Calibration Service Providers.

    The Test Bench, with 2 Sizes of Bell Prover (20L & 100L)will effectively cover Very Low (Qmin – typically 0.016 m3/h (0.27 LPM)) to Maximum (Qmax – 6 m3/h (100 LPM)) Flow Rates. Available Accuracies : 0.2% & 0.5% F.S.

    Upto 4 Gas Meters can be connected (inserted) during a campaign. Facility provided for continuously monitoring Differential Pressures (DP) across each Meter as well as upstream & downstream Line Pressures & Line Temperatures. Line Pressures are regulated typically around 500Pa.

    All the testing procedures are controlled by the PC and run automatically, the Test Results can be saved, printed.

    Simple/Basic, Manual/Semi-Automatic Test Bench Models are also available. Test Bench frame is made of Industrial Aluminium with perfect outlook and strong structure.

     
    GASCAL Series

    A Boon for Calibration of Domestic Gas Meters – OEMs, Hotels, Hospitals & R&D Institutions.

    Flow Rates : 0.018 to 6 Cu.M/Hour

    Suitable for Gas Meter Sizes : G1.6, G2.5, G4, G6

    Bell Prover : Capacity 20/100/200 Litres

    Accuracy : ±0.2% & ±0.5% (Traceable to ISO/IEC 17025 Accredited Laboratory).

    Upstream/Downstream Pressure & Temperature Measurement

    D.P. Gauges (for measurement of pressure loss) : 0 to 1.0 Inch of Water (0 to 250 Pa)

    Meter Clamping : Pneumatic

    Simultaneous Calibration of multiple Meters (Provision to mount 2/4/6/8 Meters)

    Consult us for higher capacity Bell Provers for Calibration of Industrial / Bulk Meters.


    Share your requirements for a quick response!
    Tell us what you need?

    By clicking Get Quotes Now, I accept the T&C and Privacy Policy.

    Looking for Gas Meter Calibration Test Bench?

    Quantity
    Seller Contact Details
    Waiting for permission
    To search by voice, go to your browser settings and allow access to microphone

    Allow microphone access to search with voice