Technical specification OPTICS:a. Apochromatic Optics with five step magnification changer.b. Magnification Range 2.4x-40x or morec. Max Field of view should be more than 180mm.OBJECTIVE:a.....
OPTICS: a. Apochromatic Optics with five step magnification changer. b. Magnification Range 2.4x-40x or more c. Max Field of view should be more than 180mm. OBJECTIVE: a. Working distance f=250 mm and 400 mm. b. Fine focusing for 300 mm. BINOCULAR TUBE: a. Inclinable binocular tube f=170 mm or higher. b. Inter pupillary distance adjustment from 55 mm to 75 mm. EYEPIECES: a. Pair of wide field push-in eyepieces 10x or 12.5x with rubber sleeves b. diopter setting -5D to +5D. ILLUMINATION: a. Coaxial fiber optic / direct Xenon or LED illumination system. b. In LED equipped systems: single lamp. c. Dual lamp system with quick change over in case of Xenon. d. Continuously lamp intensity adjustment by control knob near to the surgeon. STAND: a. Stable and sturdy floor stand on four lockable castors, column (height: minimum 1.6m). b. Spring-balance articulated arm, carrier arm, power supply unit. OTHERS: a. With lateral Handgrips for all knobs, dust cover. b. integrated full HD (1080p) camera & full HD recording system. Camera should have play back function for videos & photo c. Microscope is CE Marked & ISO Certified