Contact SupplierWe offer the best quality Conductivity Analyzer in the market. Our Water Conductivity Analyzer is highly advanced and is used to give measurement of contamination of water / steam with any kind of salts. We are one of the chief Steam Conductivity Analyzer Manufacturers in Pune, Maharashtra.
Benefits
Calibration methods to optimize sensor responseFully programmable cell constantTemperature measurement with Pt 100 / Pt 100Non-linear for ultra pure waterFreely programmable for concentration
For the Hach Ultra 9125 a variety of display units for choice: conductivity (S, μS, mS, per m or cm), resistivity (Ω, k Ω, mΩ, Ω per m or cm) and concentration (% or g/l)
2 calibration methods to optimize sensor response -1 or 2 points, electrical or processFully programmable cell constantPre-programmed for HCl and NaClFixed programmable coefficient in %/°C or in %/°FNon temperature compensation to conform to USP24 pharmaceutical regulations9125 offers a service menu with sensor polarization test and cable capacity measurementTotal galvanic isolation between sensors and the transmitter, between the micro-processors and other circuitry boards (analog outputs) to avoid electrical interferencesHACH ULTRA 9125 has pre-programmed concentration curves for HCl,HNO3, H2SO4, NaOH and NaClBuilt-in real time clock allows perfect data retrieval for quality control management and traceability: date, values of last calibration and self-diagnosis data
Applications
To monitor the presence or concentrations of salts in all type of waterDemineralization processTo detect the loss of efficiency of the resins (anion & cation)To monitor the quality of the ultrapure water produced right at the final outlet of the mixed bedPure waterSpecific non linear algorithms for temperature compensation in such water qualityConductivity is either measured directly (specific conductivity) or after cation exchanger (cationic conductivity)Reverse osmosis processto decide regeneration of cation / anion / mix bed and to maintain high process efficiencyTo monitor the concentrations of binary solutions of a chemical and water such as regeneration solutions for demineralization resins
Specification No. Poly / 9125 & FM 8310/11/12 | Datasheet Conductivity Analyzer | SH 1 of 1 |
---|---|---|
1.0 | General | |
1.1 | Manufacturers | HACH Ultra, Geneva, Switzerland |
1.2 | Model Number | Transmitter : Model 9125 Probe : FM 8310/ FM 8311 /FM8312 |
1.3 | Tag No. | |
1.4 | Quantity | |
1.5 | Instrument Range | Cell Constant : Conductivity Range 0.01 : 0.01 µS/cm to 200 µS/cm 0.1 : 0.1 µS/cm to 20 µS/cm 1 : 1 µS/cm to 20 µS/cm |
2.0 | Sensor | |
2.1 | Type | 2 Electrode Kohlrausch Cell |
2.2 | Electrode Material | SS 316 L |
2.3 | Connection | IP 65 Glass Polyester Connector with Low Noise Cable, with 5 Mtrs Cable |
2.4 | Temperature Rating | T max= 125o |
2.5 | Pressure Rating | P max = 10 bar |
2.6 | Temperature Sensor | Built in Pt 100 |
2.7 | Cell Constant | Select as per range mentioned above |
3.0 | Transmitter | |
3.1 | Type | Microprocessor based |
3.2 | Display Type | 4 digit, 34 x 67.4 mm backlit LCD Matrix |
3.3 | Display Details | Conductivity, Sample temperature, Alarm Status, Diagnostics, Calibration Constants, Current Outputs |
3.4 | Mounting | Panel / wall / pipe |
3.5 | Output | 2 x 0/4-20mA, DC galvanically, isolated |
3.6 | Calibration | 1 / 2 point - chemical / electrical |
3.7 | Temp. Compensation | Auto/manual |
3.8 | Temp. Compensation Type | Linear, Programmable Coefficient, Non Linear - Ultra pure, HCl, NaCl |
3.9 | Enclosure Material | Polyester coated Aluminium |
3.10 | Enclosure Protection | IP65 / NEMA 4 |
3.11 | Dimensions | 144 x 144 x 150 mm |
3.12 | Power Supply | 24 V DC |
3.13 | Diagnostics | Sensor open/short ckt, conductivity error, polarization high Offset/slope shift, temperature drift |
3.14 | Electromagnetic Compatibility | As per European standards |
4.0 | Sensor Holder Details | |
4.1 | Type | Flow through type / dip type |
4.2 | MOC | PP / SS |