Contact SupplierMeasuring System | Photometric |
Linear Measurement Range | 0.000 to 3.0000 AbsorbtionUnits (A) |
Photometric Accuracy | ±2% or 0.007 whichever is higher.from 0 to 1.5 A , ±3% from 1.5A to 3.0A |
Drift | <0.007 A/hr |
Photometric Linearity | 2.2A |
Optical Measurement | Photodiode |
Filers: | |
Type | Interference Filter |
Wavelength | 340,405,510,546, 578 & 630 nm & Two Optional |
Half Bandwidth | 10 nm ± 2nm |
Selection | Automatic by Stepper Motor |
Cuvette Volume | 500 µl |
Temperature of Cuvette Block | 370C, Peltier Controlled |
Light Source | xenon gas filled / Tungsten Halogen |
Warm upTime | 90 Sec |
Display | Four Line LCD , backlit , 16 characters |
Printer Interface | Optional Parallel Report |
Memory | 8 KB Non-Volatile RAM , Battery back up supports 40 open tests |
Analysis Mode | Absorbance : Fixed Time |
End Point : Kinetic | |
Multi -Standard :Differential | |
Ratio : PTT | |
Concentration Calculation | By factor or By Standard |
RS232 Serial Port | 2400 baud , 1 start , 8 data , 1 stop ,no parity bits |
Power | Wattage: 50 watts |
Voltage : 115-230 Volts ± 10% , 50/60 HZ | |
Operating Positions | on horizontal flat , rigid & vibration Free space |
Operating Conditions | Temperature: From + 18oC to 35oC |
Relative Humidity: up to 85% | |
Storage Conditions | Temperature: From -10oC to 60oC |
Relative Humidity: up to 85% | |
Enclosures | Abs Fire Retardant |
Size (cms) | 30 x 38 x 13.5 (lxbxh) |
Weight | 5 Kgs. |