Products / Services
  • Products / Services
  • Companies
  • Buy Leads
    Post Buy Requirement

    Bench Top Tds Meter

    • EC meterConductivity meter
    • TDS meternacl meter
    • Supply TypeManufacturer, Exporter, Supplier, Retailer
    • Preferred Buyer Location All over the world

    We are offering bench top tds meter. multi-ranging ec, tds, nacl, temperature meter hi 2300 multi-ranging ec, tds, nacl, temperature meter with 4-ring potentiometric conductivity probe with....
    View More Details
    Send Enquiry

    Company Information

    • calendar Member Since 19 Years
    • building Nature of Business Retailer
    • Year of Establishment 2005

    Ask for more detail from the seller

    Contact Supplier
    Report incorrect details
    • Product Details

    • Company Details

    no_img_icon

    We are offering bench top tds meter.
    multi-ranging ec, tds, nacl, temperature meter hi 2300
    multi-ranging ec, tds, nacl, temperature meter with 4-ring potentiometric conductivity probe with platinum sensors hi 2300
    features:
    auto-ranging
    4-ring potentiometric conductivity probe with internal temperature sensor
    automatic (atc), manual (mtc) or no temperature compensation (notc)
    good laboratory practice
    pc compatible
    specifications:
    range: (ec) 0.00 to 29.99 µs/cm; 30.0 to 299.9 µs/cm; 300 to 2999 µs/cm; 3.00 to 29.99 ms/cm; 30.0 to 200.0 ms/cm; up to 500.0 ms/cm (actual ec)*
    range: (tds) 0.00 to 14.99 mg/l (ppm); 15.0 to 149.9 mg/l (ppm); 150 to 1499 mg/l (ppm); 1.50 to 14.99 g/l (ppt); 15.0 to 100.0 g/l (ppt); up to 400.0 g/l (actual tds)*, with 0.80 conversion factor
    range: (nacl) 0.0 to 400.0%
    range: (temp) -9.9 to 120.0°c
    resolution: (ec) 0.01 µs/cm; 0.1 µs/cm; 1 µs/cm; 0.01 ms/cm; 0.1 ms/cm
    resolution: (tds) 0.01 mg/l; 0.1 mg/l; 1 mg/l; 0.01 g/l; 0.1 g/l
    resolution: (nacl) 0.1%
    resolution: (temp) 0.1°c
    accuracy: (ec) ±1% of reading ± (0.05 µs/cm or 1 digit)
    accuracy: (tds) ±1% of reading ± (0.03 mg/l or 1 digit)
    accuracy: (nacl) ±1% of reading
    accuracy: (temp) ±0.4°c
    calibration: (ec) automatic, 1 point with 6 memorized values (84, 1413, 5000, 12880, 80000, 111800 µs/cm)
    calibration: (nacl) 1 point, with hi 7037 calibration solution
    calibration: (temp) 2-point, at 0 and 50°c (32 and 122°f)
    temperature compensation: automatic or manual, 0 to 60°c (32 and 140°f)
    temperature coefficient: selectable from 0.00 to 6.00%/°c (ec and tds only)
    tds conversion factor: selectable from 0.40 to 0.80 (default value: 0.50)
    probe: (included) hi 76310, 4-ring platinum, internal temperature sensor
    auto-off: after 5 minutes of non-use (can be disabled)
    power supply: (included) 12 vdc adapter
    environment: 0 to 50°c (32 to 122°f); rh max 95%
    dimensions/weight: 240 x 182 x 74 mm / 1.1 kg
    product contents: hi 2300 is supplied with conductivity probe, power adapter and instructions.


    Share your requirements for a quick response!
    Tell us what you need?

    By clicking Get Quotes Now, I accept the T&C and Privacy Policy.

    Looking for Bench Top Tds Meter?

    Quantity
    Seller Contact Details
    Waiting for permission
    To search by voice, go to your browser settings and allow access to microphone

    Allow microphone access to search with voice