Contact SupplierSpecification of 21l
horizontal autoclave with an internal volume of 21 l , built in steel and aluminum chemically treated against corrosion sterilization chamber , external finishing with epoxy paint cured at 200 ° c with maximum resistance to temperature and oxidation, 1 cycle automatic sterilizing, manual filling 400 ml , safety valve against overpressure , overtemperature safety thermostat , door latch lever type that prevents the door from opening with the airlock door seal system with special silicone rubber, safety fuse in case of voltage variation and 3 trays for surgical instruments . Warm-up time 30 minutes; sterilization time 10 minutes , drying time 30 minutes
pressure sterilization : 2.3 kgf cm ² + - 0.2 kgf cm ² .
Working pressure : 1.4 to 2.5 kgf cm ³
operating temperature: 134 ° c .
Width : 375mm
height : 360mm
power : 1200 w
weight: 46 kg
voltage : 220 v or 110 v
consumption : 0.651.2 kw
note: ohmic value resistance
- resistors 127v x 12 ω.
- resistance 220v x 42 ω.
Type of resistence: external strap
power: (important)
- fuse protection internal plate = 2
- 220v-8a-1200w (10a fuse)
- 110127v- 14a-1200w (15a fuse)
- frequency = 5060hz
- voltage range = ± 2% (in tests conducted with a range of up to 10%, there was normal).
Ground cable
- make sure that the ground wire installed on the outlet is actually connected to "earth". Some electricians install the ground wire in the neutral (which is incorrect)