Listing ID #725788
Company Information
Ask for more detail from the seller
Contact SupplierWe offer Medical Waste Sterilizer.
1. for gravity flow autoclave
• The temperature of not less than 121°C and pressure of not less than 1.02 K.G.CM2 and autoclave residence time not less than 60 minutes.
• The temperature of not less 135°C and pressure of not less than 2.1 K.G.CM2 and autoclave residence time not less than 45 minute.
• The temperature of not less than 149°C and pressure of not less than 3.5 K.G.CM2 and autoclave residence time not less than 30 minute.
2. for vacuum autoclaving with pre vac & purge system.
• The temperature of not less than 121°C and pressure of not less than 1.02 K.G.CM2 and autoclave residence time not less than 45 minutes with one or more pre vacuum pulse.
• The temperature of not less than 135°C and pressure of not less than 2.1 K.G.CM2 and autoclave residence time not less than 30 minutes with one or more pre vacuum pulse.
Construction:–
Shape: – Either cylindricalrectangular. Chamber is made of boiler grade sheetAISI 304AISI 316 thick sheet having one or two nos. pivoted door mounted on a sturdy structural stand dully insulated & covered either by S.S. or M.S. Sheet. Pit mounted design is also available as per the customer requirement for easy loading & unloading. The door will be provided with automatic safety door looking unlocking mechanism. Valves are provided for operation safety steam inlet, drainage etc.
Control and Monitoring System:–
Microprocessor based computer system and printer is provided for controlling and monitoring the sterilizing cycle with the following feature
• Fully programmable microprocessor based unit with no volatile memory and with different program combinations.
• Automatic operation of the sterilizer.
• Self diagnostic system gives fault massage to the operator.