Our Product / Services
Advanced Micro-controller based TechnologyHeavy Duty Portable / Bench-top ModelHigh Accuracy, Resolution, Uniformity & StabilityLarge Graphic Display (Multi-information Display)Measurement Capability for TC, RTD, mV / mA / V & Resistance SignalsTouch Membrane KeypadSwitch Test Contact MeasurementMulti-hole Thermowell InsertsRS-232 Interface & Calibration SoftwareCount down timer (Shows when unit is stable)Stability Indicator
Type | Range | Resolution | Accuracy (% of reading) | Type | Range | Resolution | Accuracy | |
mV (Low) | 0 to 100 | 0.001 | ±(0.025%+2µV) | TC Type R | 150 to 1700 | 0.1 | ±(0.025%+0.2°C) | |
mV (High) | 0 to 1000 | 0.01 | ±(0.025%+10µV) | TC Type S | 170 to 1700 | 0.1 | ±(0.025%+0.2°C) | |
Volts | 0 to 10 | 0.0001 | ±(0.025%+0.08mV) | TC Type N | 0 to 1300 | 0.1 | ±(0.025%+0.1°C) | |
mA | 0 to 25 | 0.0001 | ±(0.025%+0.4µA) | RTD Pt100 | -100 to 800 | 0.01 | ±(0.025%+0.1°C) | |
Ohms | 0 to 500 | 0.001 | ±(0.025%+12mâ¦) | RTD Pt200 | -100 to 600 | 0.1 | ±(0.025%+0.15°C) | |
KOhms | 0 to 3.5 | 0.00001 | ±(0.025%+120mâ¦) | RTD Pt500 | -100 to 530 | 0.1 | ±(0.025%+0.1°C) | |
TC Type J | -100 to 1200 | 0.1 | ±(0.025%+0.1°C) | RTD Pt1000 | -100 to 650 | 0.01 | ±(0.025%+0.1°C) | |
TC Type K | -60 to 1260 | 0.1 | ±(0.025%+0.1°C) |
Compact, Light Weight & Portable
High Emissivity, Wide Aperture
Excellent Accuracy & Stability
Auto-tune PID Controller
Low Power Consumption
Cost effective Best value for Money
RS-232 Interface & Calibration Software Optional
2 Line Alpha Numeric Display
Touch Membrane Keypad
Selectable Engineering Pressure Units
DC Current Measurement : 0 to 25 mA (Active)
Switch Test capability & Zero Offset facility
Audio & Visual Alarms for Over Pressure
Pressure Transducer all Media compatible with SS316
9V Battery operated with Low Battery indication
Digital Pressure Indicator with matching Pneumatic Hand Pump with Vernier
Compact, Light Weight Unit with Holster
Selectable Temperature Output Readings (in C / F)
Measurement & Simulation of
Thermocouple types R, S, B, E, K, J, T & N
Voltage 1 V Simulation, 100 mV Measurement
Input | Range | Input range | Resolution | Accuracy | Remarks |
Voltage (DC) | 100 mV | -10 to 110 mV | 10 mV | ±0.05% measured value
± 30 mV |
Input Resistance 1MW |
Thermocouple | R | -40 to 1760°C | 1°C | ±0.05% measured value ±3°C (<100°C)
±0.05% measured value ±2°C (>100°C) |
Input Resistance 1MW |
S | -20 to 1760°C | 1°C | |||
B | 400 to 1800°C | 1°C | ±0.05% measured value ±3°C (400 to 600°C)
±0.05% measured value ±2°C (>600°C) |
||
E | -200 to 1000°C | 0.1°C | ±0.05% measured value ±2°C (< -100°C)
±0.05% measured value ±1°C (> -100°C) |
||
K | -200 to 1370°C | 0.1°C | |||
J | -200 to 1200°C | 0.1°C | |||
T | -200 to 400°C | 0.1°C | |||
N | -200 to 1300°C | 0.1°C |